Products

Gri3D® Hydrogel Microcavity Plate - Plastic

InSphero Gri3D hydrogel microcavity plate: PEG hydrogel U-shaped microcavities, 31–121 per well, giving uniform organoids in one imaging plane for high-throughput organoid culture, imaging and screening.

Product information

Brand
InSphero
Catalog no.
GRI3D-96P-S-96
Specification
10/pack

Product overview

In 2025 InSphero announced the acquisition of DOPPL SA, gaining the Gri3D organoid culture technology behind Sun Bioscience. The aim was to combine InSphero's established 3D spheroid and organ-on-chip platform with Gri3D's highly consistent organoid culture to form a more complete 3D in vitro model system.

The Gri3D hydrogel microcavity plate is a ready-to-use 3D cell-culture plate based on PEG hydrogel microcavity technology for high-throughput, standardised and reproducible organoid and spheroid culture. Its advantages:

1. Highly consistent organoid formation

Pre-formed and ready to use, with cell-repellent PEG hydrogel U-shaped microcavities that let cells aggregate naturally and reliably form uniform, size-controlled organoids and spheroids, from tens to thousands of cells.

2. Built for high-throughput, standardised experiments

Based on a standard 96-well plate (clear or black), each well contains an array of 31–121 microcavities. The bottom can be polystyrene (for biochemical or lysis assays) or COP (for high-content imaging), allowing flexible experimental design. The plate complies with ANSI/SLAS and SBS standards and connects seamlessly to automated liquid handling and high-content screening systems for large-scale work. Microcavity diameters range from 400 to 800 µm.

3. Excellent optical imaging

3D imaging becomes easier. The hydrogel contains more than 95% water and is highly transparent, reducing light refraction and image distortion; all organoids sit in one imaging plane, suiting high-resolution confocal imaging and quantitative high-content analysis.

4. Convenient, low-loss liquid handling

A dedicated pipetting port protects organoids during medium exchange, avoiding tissue loss and supporting long-term culture and tracking of individual organoids, with manual or automated handling.

5. Ready-to-use design that lowers the barrier to 3D culture

No complex hydrogel preparation or specialist handling is needed: researchers simply seed, culture and analyse, which speeds up 3D organoid experiments and improves reproducibility.

Frequently asked questions

Product and storage

1. What is Gri3D, and what is it made of?

A 3D culture plate based on PEG hydrogel microcavities that forms relatively uniform organoids or spheroids in the microcavities for imaging, drug screening and many downstream analyses. The culture plate is polystyrene, the imaging plate COP, and the microcavities are cell-inert PEG hydrogel, providing a matrix-free 3D environment for cell aggregation and organoid formation.

2. How should it be stored, and for how long?

Store at 4 °C on arrival; unopened product is stable for at least 6 months at 4 °C.

3. Which microcavity sizes are available, and how do I choose?

Microcavity diameters are about 400–800 µm. Choose small microcavities for smaller organoids or fewer starting cells – higher microcavity density and more data points; choose large microcavities for long-term culture or larger organoids.

4. What volumes are recommended? Can it be automated?

Per well, 50 µL in the cell-seeding chamber and 150 µL in the pipetting port, 200 µL working volume in total. Hydrogel equilibration, cell seeding, medium exchange and compound treatment can all be combined with liquid-handling workstations and other automation.

Cell culture

5. Which organoids has it been used for?

Intestinal, liver, colorectal cancer, pancreatic cancer and blood–brain barrier organoids, among others; applications need optimising for the cell type and conditions.

6. How is it seeded? How uniform is it? Is centrifugation or coating needed?

Seed each well with 50 µL of single-cell suspension and control the final organoid size through the starting density. A specially designed seeding ring reduces the meniscus effect so cells fall more evenly into the microcavities; the variation in cell number between microcavities in one well is usually below 5%. Seeding and aggregation take a single pipetting step without centrifugation, and because PEG hydrogel repels cells, no extra coating is needed.

7. How long until organoids form, and how long can they be cultured?

Cells usually aggregate fully into compact structures in 2–5 days, depending on cell type, proliferation rate and starting density. Long-term culture is possible as needed – for example, embryonic-stem-cell-derived retinal organoids for 26 days and primary human hepatocyte spheroids for 21 days.

8. Is co-culture possible?

Yes: different cells can be mixed and seeded together, or a second cell type (such as immune cells) added after organoids form. Applications include blood–brain barrier organoid co-culture and T-cell killing assays.

Medium and ECM

9. How do I exchange medium without harming the organoids, and how often?

Always work through the pipetting port: remove 150 µL from the port, then add the same volume of fresh medium. Exchange usually every 2–3 days, or more often if needed.

10. Is ECM needed, and which are compatible?

It depends on the model. For organoids that expand in basement-membrane extract, ECM can be mixed into the medium – for example, Matrigel is usually added at about 1.5–2% at each medium exchange. Collagen I, Matrigel, growth-factor-reduced Matrigel, Cultrex BME, laminin and other matrices that can be diluted in medium are compatible, with concentrations optimised for the model.

11. Can unused wells be used later? Can the plate be washed and reused?

Remaining wells can be kept for other experiments, but unused microcavities must stay hydrated and sterile; check hydration before use. Washing and reuse are not recommended, as the manufacturer cannot guarantee quality on reuse.

12. How do I avoid damaging the hydrogel?

Always exchange medium and add compounds through the pipetting port; when seeding, let the tip touch the rigid plastic seeding ring, not the hydrogel. PEG hydrogel is delicate, and direct disturbance by a tip can destroy the microcavity structure and lose organoids.

Assays and downstream analysis

13. Which assays can be run directly in the plate?

Imaging, immunostaining, supernatant assays, retrieval of individual organoids and more. All immunofluorescence steps – fixation, permeabilisation, blocking, antibody incubation and imaging – can be done in Gri3D; transmitted-light, fluorescence and confocal imaging are supported, and organoids sit in a relatively uniform focal plane so many can be imaged at once.

14. Is it suitable for high-throughput screening?

Yes. The SLAS-standard 96-well format, multiple organoids per well and a uniform focal plane make it easy to set up image-based organoid and drug screens.

15. Are fluorescence, colorimetric and viability assays possible?

Fluorescence and colorimetric (O.D.) assays are possible; because the plate is clear, the manufacturer recommends transferring samples to white or black plates for better results. Viability can be measured by fluorescence imaging such as LIVE/DEAD or by luminescence such as CellTiter-Glo 3D; the CellTiter-Glo 3D reagent does not act on the PEG hydrogel and does not interfere with the luminescence readout.

16. Can supernatant be sampled and organoids harvested?

Culture medium can be collected safely through the pipetting port for mass spectrometry and other analyses. Organoids grow in open microcavities without a solid matrix, so simple pipetting harvests them all or picks individual microcavities, with no complex matrix removal; the small amount of PEG carried over is cell-inert and removed by centrifugation. Harvested organoids can be passaged, or DNA, RNA and protein extracted by standard methods, pooling wells if the amount is small.

17. Can cryosections be prepared?

Yes: fix and cryopreserve in the microcavities, remove the hydrogel containing the organoids, embed in OCT, snap-freeze and section.

18. Does PEG hydrogel absorb small-molecule drugs?

In most cases there is no marked binding, but PEG may bind thiol-containing molecules; for such compounds, assess according to their molecular properties.