Akura™ 96 Spheroid Microplate

InSphero Akura 96 ultra-low-attachment spheroid microplate: SureXchange ledge for safe medium exchange, COP plate with 0.8 mm well bottom for high-content imaging, 70 µL per well, 250 RCF for 2 min, spheroids usually in 2–5 days.

Product information

Brand
InSphero
Catalog no.
CS-PB13 / CS-PB15
Specification
96-well spheroid microplate (5 plates/pack, 20 plates/pack)

Product overview

InSphero Akura 96 and Akura 384 ultra-low-attachment microplates are a simple, flexible platform compatible with high-throughput and automated systems for scaffold-free 3D cell culture, observation and assessment. Each plate consists of ultra-low-attachment culture wells and a lid that reduces medium evaporation. They suit cell types known to work in 3D culture – immortalised or modified cell lines, primary cells and induced pluripotent stem cells (iPS cells) – for forming organoids, spheroids or microtissues. The bottom of the Akura 384 plate is a gas-permeable membrane that ensures sufficient gas exchange during long-term culture.

Design

From a distance an Akura 96/384 plate looks like an ordinary microplate: its dimensions follow the SLAS/ANSI standard and it fits routine handling and instruments. Up close, each well has an "inverted chimney" shape and a pipette-tip docking point called the SureXchange ledge for medium exchange. In the Akura 96 plate the ledge is axially symmetric; the Akura 384 plate has a ledge on one side only. When exchanging medium, rest the pipette tip on the ledge for tactile feedback that stops the tip going too deep and disturbing the spheroid. The well walls carry a durable coating that prevents spheroids from attaching during culture.

The Akura 384 plate has 24 columns (1–24) and 16 rows (A–P); well A1 is at the top left and P1 at the bottom left, which defines the plate orientation. Its wells are asymmetric: the ledge sits towards the top of the plate and the cavity in which the spheroid grows towards the bottom, so pay attention to orientation when using the Akura 384 plate.

Advantages

  • Convenient spheroid formation by cell self-aggregation in ultra-low-attachment wells.
  • The well design makes medium exchange easier and prevents loss of or damage to spheroids during long-term culture.
  • A 1 mm flat-bottom viewing window at the bottom of each well makes spheroids easy to locate and observe, avoiding the local distortion of round-bottom ULA plates.
  • The Akura 384 plate is compatible with advanced imaging and automated liquid-handling systems for high-throughput work; its black well walls eliminate fluorescence crosstalk between wells.

Protocol summary (Akura 96)

  • Prepare: cells and suitable medium, inverted microscope, cell counter, pipettes or automated liquid handler, plate centrifuge, CO₂ incubator (5% CO₂, 95% relative humidity, 37 °C). Pre-warm the medium to 37 °C, wipe the plate packaging with 70% ethanol and open it under sterile conditions.
  • Pre-wet: add 40 µL of medium containing FCS or BSA per well, with the tip close to but not touching the well bottom (an 8- or 12-channel pipette saves time), pipette gently to wet the well without bubbles, then rest the tip on the ledge and remove all medium. Less than 5–7 µL remains in an Akura 96 well (less than 2–3 µL in an Akura 384 well), which can be ignored.
  • Seed: prepare a single-cell suspension, count the cells, mix thoroughly and make up the chosen number of cells to 70 µL (Akura 96) or 50 µL (Akura 384). For long-term culture of proliferating cells use a low starting density of 250–500 cells per well; for non-proliferating cells or to obtain larger spheroids quickly, 2,500–25,000 cells per well. Dispense slowly (below 10 µL/s) without touching the well bottom.
  • Aggregate: optimise conditions where needed (seeding density, medium, supporting cells such as matrix-secreting cells, or additives such as ECM); for new cell types try several conditions. Centrifuge the lidded plate at 250 ×g (RCF) for 2 min – optional, but it removes bubbles and helps cells settle to the well bottom. Place the plate in the incubator tilted at 30° or on the InSphero Akura tilting stand (CS-10-002-00) and culture for 2–5 days, checking spheroid morphology daily.
  • Exchange medium: usually 2–3 times a week. Slide the tip down the well wall until you feel the ledge, aspirate slowly (below 10–20 µL/s; up to 50 µL/s is acceptable when adding to Akura 96 wells), then add 70 µL (Akura 96) or 50 µL (Akura 384) of pre-warmed medium the same way. Repeat for a more complete exchange.
  • Harvest from Akura 96: pre-wet the tip with at least 60 µL of 100% FCS so spheroids do not stick. Either use a 1000 µL tip, which is slightly wider than the well bottom and sits just above the spheroid at the ledge, or a 100–200 µL tip held at a small angle along the wall to the well bottom.

Frequently asked questions

1. What has improved in the new-generation Akura 96?

  • COP material with about 92% light transmission at 400–800 nm;
  • well-bottom thickness reduced from 1.3 mm to 0.8 mm;
  • skirt height reduced to 0.4 mm, supporting high-numerical-aperture objectives;
  • plate flatness below 80 µm, which helps automation;
  • a custom low-evaporation lid about 200 µm from the plate, helping reduce edge-well evaporation;
  • standard SLAS plate height of 14.35 mm;
  • maximum well volume increased from 170 µL to 280 µL.

2. Why pre-wet the plate before use?

To reduce bubbles during seeding. Add 40 µL of medium per well, then remove it completely; a small amount of liquid may remain at the bottom (usually less than 5–7 µL).

3. How many cells should be seeded per well, and in what volume?

For long-term growth studies start from 250–500 cells per 70 µL per well; for non-proliferating cells or to form larger spheroids quickly, start from ≥2,500 cells per 70 µL per well. Use 70 µL of cell suspension per well, dispensed slowly at less than 10 µL per second with the tip close to but not touching the well bottom; mix gently before seeding for more uniform spheroid size and composition. Test several densities for a new cell type.

4. Why centrifuge after seeding?

A short spin removes bubbles and helps cells settle and aggregate into spheroids. Put the lid on, centrifuge in a plate centrifuge at 250 RCF for 2 minutes, then place the plate in a humidified 37 °C CO₂ incubator; spheroids usually form in 2–5 days.

5. Will every cell type form spheroids?

Not guaranteed. Spheroid formation depends on the tissue of origin, cell morphology and function, and the isolation process; the same cell type can also differ between donors. For new cell types, optimise the cell concentration, medium composition, supporting cells or ECM.

6. How do I exchange medium without aspirating the spheroid?

Each well has a SureXchange ledge as a pipette-tip locating point: place the tip on the ledge and aspirate slowly, which reduces the risk of removing the spheroid; about 5–7 µL usually remains in the well. Then add 70 µL of fresh medium from the ledge in the same way at less than 50 µL per second. With automated liquid handlers, a similar effect can be achieved by adjusting the vertical tip position.

7. How can edge-well evaporation be reduced? Is the plate suitable for long-term culture?

Evaporation from outer wells is common in low-volume systems; when severe, it can cause salts, serum and other components to concentrate or even precipitate, affecting spheroid condition or changing the actual drug concentration. Recommendations: use an incubator with relative humidity >95% and open the door as little as possible; keep 50–70 µL of medium per well, increasing to at most about 80 µL if humidity control is poor; exchange medium more often as needed (for example every two days or daily); for long-term culture with infrequent exchanges, use the InSphero Incubox to reduce edge effects. With humidity and evaporation managed, Akura 96 can be used for long-term spheroid culture.

8. Can it be used for high-content imaging? What should I watch for?

Yes. The plate is optimised for imaging and works with many high-content imaging systems. Note: the well bottom is 0.8 mm thick and non-continuous, so some systems need a new 96-well plate definition and may need a wider autofocus range; for Z imaging of a whole spheroid the objective working distance must cover 0.8 mm plus the sample height; objectives with a correction collar should be set to 0.8 mm.