Products
Akura™ PLUS Hanging Drop System
InSphero Akura PLUS Hanging Drop System: 40 µL drops, spheroid diameter usually within ≤5% across a plate, transferred into Akura 96 with 70 µL for long-term culture and analysis.
Product information
- Brand
- InSphero
- Catalog no.
- CS-PF24
- Specification
- 96-well hanging-drop culture system (10 sets/pack)
Product overview
The InSphero Akura PLUS Hanging Drop System is a simple, flexible, automation-compatible platform for scaffold-free 3D cell culture, observation and assessment. Each system consists of one Akura PLUS hanging-drop plate and one InSphero Akura 96 microplate. InSphero itself uses this system for large-scale scaffold-free 3D culture, organoid culture and production of ready-to-use microtissues.
Advantages
- The Akura PLUS hanging-drop plate forms robust drops containing spheroids, making scaffold-free 3D culture and organoid culture straightforward.
- Drops containing spheroids are easily transferred into the InSphero Akura 96 microplate for further culture.
- In the specially designed Akura 96 microplate, spheroids can be cultured, analysed and observed over long periods.
Protocol summary
- Prepare: cells and suitable medium, inverted microscope, cell counter, pipettes or automated liquid handler, plate centrifuge, CO₂ incubator (5% CO₂, 95% relative humidity, 37 °C) and PBS. Pre-warm the medium to 37 °C, wipe the packaging with 70% ethanol, open it under sterile conditions and check the components: the Akura PLUS hanging-drop plate (reservoir base plate, hanging-drop plate frame, lid, humidification pad and tweezers) and the Akura 96 microplate (plate and lid).
- Humidification pad: put 20 mL of 0.5× PBS into a sterile container such as a 10 cm dish, soak the pad with tweezers until fully wet (about 5 min), lift the frame out of the base plate, place the pad in the base plate and replace the frame.
- Seeding and drop formation: prepare a single-cell suspension, count, mix thoroughly and make up the chosen number of cells to 40 µL. For long-term culture of proliferating cells seed 250–500 cells per well; for non-proliferating cells or to obtain larger spheroids quickly, 2,500–25,000 cells per well. Dispense the 40 µL slowly with the tip in close contact with the plate. Culture with the lid on in the CO₂ incubator; most cell types form compact spheroids after about 4 days. Move the hanging-drop plate with care so the drops do not fall.
- Transfer for long-term culture: pre-wet the Akura 96 microplate first. Lift the hanging-drop frame and place it on the Akura 96 plate, aligned by the three pins around the frame. Slowly add 70 µL of medium to each well of the frame with the tip in close contact with the plate; the drop falls into the corresponding Akura 96 well. Check transfer under an inverted microscope, centrifuge the lidded Akura 96 plate at 250 ×g (RCF) for 2 min to remove bubbles and settle the cells, then exchange the medium (remove all and add 70 µL of fresh medium) so every well holds an accurate volume.
Key parameters
| Item | Akura PLUS Hanging Drop System |
|---|---|
| Culture mode | Hanging drop |
| Plate format | 96-well |
| Recommended drop volume | 40 µL per drop |
| Cells per drop for long-term culture | 250–500 |
| Large spheroids / fast formation | ≥2,500 per drop |
| Spheroid size uniformity | Diameter variation within one plate usually ≤5% |
| Spheroid transfer | To Akura 96 Plate by adding 70 µL |
| Akura 96 culture volume | 50–70 µL per well |
| Recommended incubator humidity | >95% RH |
| Observation | Inverted bright-field microscope, 10× long-working-distance objective |
Frequently asked questions
1. What is Akura PLUS?
A hanging-drop system for culturing 3D spheroids and microtissues, used for spheroid formation, long-term culture, drug treatment and subsequent analysis.
2. How many cells, and what volume, per drop?
Use 40 µL of cell suspension per drop, dispensed at <50 µL per second for a stable drop, and mix gently before seeding. For long-term growth studies start from 250–500 cells per 40 µL; for non-proliferating cells or to form larger spheroids quickly, start from ≥2,500 cells per 40 µL. Run a density pilot for new cell types.
3. What is SureDrop technology, and how are drops kept uniform?
SureDrop is the microstructure technology of the Akura PLUS hanging-drop plate that helps accurate dispensing and aspiration. By controlling drop volume and seeding number, spheroid diameter within one 96-well plate can usually be kept within 5%. When seeding, the tip must make full contact with the well surface; springs in the plate adjust the contact automatically during pipetting to keep the seal and liquid transfer stable.
4. How are spheroids transferred to the Akura 96 Plate?
First pre-wet the Akura 96: add 40 µL of Aggregation Medium per well and remove it completely, leaving only a little liquid at the bottom, to reduce bubbles. Then align Akura PLUS with Akura 96 using the locating pins and holes and add 70 µL of medium or buffer through the Akura PLUS inlet at ≤10 µL per second; the spheroid drops into the Akura 96 well below.
5. Why is evaporation control especially important for hanging drops, and how is it done?
Hanging drops are a low-volume system and outer wells evaporate easily, which can concentrate or even precipitate salts and serum, affect spheroid formation and cell condition, and may change the actual drug concentration in long-term studies. Recommendations: use a well-controlled incubator with humidity >95% and open the door as little as possible; during culture use the supplied humidification pad, fully wetted with sterile 0.5× PBS, or a hypotonic buffer such as 0.2× PBS if humidity control is poor; after transfer to Akura 96 keep at least 50–70 µL per well, increasing to at most about 80 µL and exchanging medium more often if needed; for long-term culture with infrequent exchanges, use the InSphero Incubox.
6. What if my cells do not form spheroids?
Possible factors include cell type and cell-line properties, harvest/pre-treatment conditions, seeding density, co-culture ratio, medium formulation, culture time and evaporation. For most tumour cell lines that aggregate easily, try Akura 96 first; for primary cells, complex co-cultures or cells that do not form spheroids even in ultra-low-attachment plates, Akura PLUS is recommended. Some cells also need co-culture with fibroblasts or specific growth factors to form compact spheroids or keep an organotypic phenotype.
7. How are spheroids observed and imaged?
Use an inverted bright-field microscope to follow formation, morphology and growth, preferably with a 10× long-working-distance objective. To view all 96 wells, transfer the hanging-drop plate onto an empty sterile base plate for imaging, then return it to the base plate with the humidification pad and back into the incubator.