Technical column
Stop trusting the average LNP size: TRPS resolves a 71 nm → 66 nm loading shift and shows subpopulations, concentration and charge particle by particle
A single DLS average blurs empty particles, aggregates and charge subpopulations into one number. IZON nanopore technology measures LNPs one particle at a time: on Exoid, empty and mRNA-loaded LNPs are clearly distinguished at modal sizes of 71 nm and 66 nm, and zeta potential appears as a full distribution. Pulsoid takes 2–5 minutes per sample with 99.7% agreement with TEM for QC release.
Definition
TRPS (tunable resistive pulse sensing) records the ionic-current blockade produced as each particle passes through a nanopore; with NIST-traceable calibration particles it yields single-particle size, true particle-number concentration and single-particle zeta potential.

TRPS records individual blockade pulses and can place size, particle concentration and zeta potential within one LNP quality framework.
The DLS average for an LNP lot can look perfectly normal while the sample actually contains empty particles, loaded particles, aggregates and free lipid or protein impurities. One average cannot answer "how many kinds of particle are in this lot, and in what proportions?" — yet that is exactly what drives biodistribution, cellular uptake, lot consistency and the credibility of regulatory data.
IZON Science's single-particle nanopore technology takes the question apart: every particle is measured and counted individually. On Exoid, empty LNPs of one formulation have a modal size of 71 nm and 66 nm after mRNA loading — a difference of a few nanometres that is clearly visible in the single-particle distribution. Zeta potential stops being one number and becomes a charge distribution in which subpopulations and drift can be seen.
What the average hides: three key LNP parameters
Lipid nanoparticles (LNPs) have become a major delivery vehicle for mRNA vaccines, siRNA drugs and gene therapy. As mRNA pipelines advance rapidly, LNP critical quality attributes (CQAs) — size, concentration, zeta potential, polydispersity and subpopulation composition — directly determine biodistribution, cellular uptake efficiency, lot consistency and the credibility of regulatory submissions. Three parameters form the core of characterization:
- Size governs tissue penetration, uptake route and in-vivo clearance rate, and is an important indicator of stability and delivery efficiency;
- Particle concentration maps directly to delivered dose and capacity calculations, and is a key number for scale-up and lot release;
- Zeta potential reflects surface charge and colloidal stability, affecting storage stability, aggregation tendency and interaction with cell membranes.
Beyond these three, sample heterogeneity is less visible but just as critical: empty and loaded particles, intact particles and aggregates, and target particles and free lipid/protein impurities often coexist in one system. The long-standard ensemble techniques such as dynamic light scattering (DLS) report a single "average size", masking the subpopulations, aggregation and lot-to-lot drift that affect developability.
How TRPS works: measure and count every particle
Tunable resistive pulse sensing (TRPS) uses a tunable nanopore. LNP suspension is driven by pressure through the pore one particle at a time, and each particle creates a transient blockade pulse in the ionic current:
- Pulse magnitude is proportional to particle volume, giving the size of every particle;
- Pulse frequency is proportional to particle flux; with extrapolated calibration across several pressure points it gives particle concentration;
- The speed at which a particle crosses the pore reflects its electrophoretic mobility, from which single-particle zeta potential is derived.

Blockade magnitude, event frequency and pressure/voltage response support interpretation of size, concentration and zeta potential.
Measurements are standardized throughout with NIST-traceable calibration particles, so data are comparable and reproducible across instruments, operators and laboratories — the measurement-quality system that regulatory submissions require.
The single-particle mechanism gives four direct advantages:
- Single-particle resolution: a size distribution is reported directly rather than inferred from an average, so highly heterogeneous samples are no longer a blind spot;
- Simultaneous size and concentration: both parameters from one run;
- True concentration: no assumptions about particle optical properties; the result is a true particle-number concentration;
- Wide dynamic range: 40 nm–11 µm in size and 10⁵–10¹¹ particles/mL in concentration, compatible with common electrolytes such as PBS, HEPES, MES and KCl.
Note that size-plus-concentration and size-plus-zeta-potential are acquired under their own method set-ups; the three readouts should not be read as one simultaneous acquisition.
Empty vs mRNA-LNP: single-particle data turn "a few nanometres" into evidence
When the average zeta potentials of empty and mRNA-loaded LNPs differ only slightly, one average value hides the real heterogeneity inside the sample: charge subpopulations cannot be distinguished and population changes caused by nucleic-acid encapsulation are hard to capture.
mRNA carries a large negative charge, and encapsulating it inside an LNP changes the particle's overall surface electrical properties. TRPS detects every particle individually, collecting single-particle size, true particle concentration and single-particle zeta potential, and outputs a complete, continuous zeta-potential histogram that:
- shows the charge shift and the degree of overlap between the empty and loaded groups;
- identifies particle subpopulations with different charge characteristics instead of relying on one pooled average;
- reveals the population-level shift of LNPs toward more negative zeta potential after nucleic-acid loading.
IZON application data show that when empty LNPs and mRNA-LNPs of the same formulation were compared on Exoid, the modal size was 71 nm for empty particles and 66 nm after mRNA loading. This difference of a few nanometres is clearly resolved in the single-particle distribution; it reflects the real effect of nucleic-acid loading on LNP structure and is direct evidence for formulation optimization and loading-efficiency assessment. Combined with the single-particle size distribution and true particle concentration, development teams can compare empty and loaded LNPs by size, concentration and charge population, giving multidimensional single-particle evidence for formulation development, process optimization and encapsulation assessment.
Similar decision points include tracking the rate and proportion of aggregate formation in storage-stability studies, comparing subpopulation consistency between lots, and assessing how upstream and downstream process changes affect size and charge distributions. All of them need single-particle data.
Zeta potential: from one average to a charge distribution
Zeta potential is the potential at the particle's slipping plane and an important parameter for assessing LNP colloidal stability and uptake potential. Conventional methods such as PALS electrophoretic light scattering give only the whole-sample average, while LNP development really needs to know: how broad is the charge distribution? Is there a low-charge subpopulation prone to aggregation? Has the charge distribution drifted after a process change?
IZON nanopore systems track the electrophoretic motion of individual particles as they cross the pore under an electric field, measuring zeta potential particle by particle together with the size of that same particle. Size and charge are paired at the single-particle level — a depth of analysis ensemble techniques cannot provide.
In measurements of multimodal mixtures (standard particles with different surface modifications), TRPS not only separated the average zeta potential of each subpopulation but directly showed the width of each subpopulation's charge distribution, whereas most techniques return a single merged average. For formulation screening, stability tracking and change control, that is the difference between seeing a risk and missing it.

Zeta potential is one interfacial readout; it does not independently prove storage stability, delivery efficiency or in-vivo potency.
Two points help interpret it correctly: zeta potential describes the potential near the slipping plane, not the direct surface charge density; and LNPs can also be stabilized by PEGylated lipids, steric hindrance and formulation structure. Compare formulations in the same medium, pH, ionic strength, temperature and instrument method; otherwise a change may come from the measurement environment rather than the LNP itself.
When the average stays quiet but the sample has changed: LNP stability, particle by particle
The change most easily missed in a stability study is not an average size that clearly fails, but an average that barely moves while a small population of aggregates has already appeared in the distribution. IZON ran a clear comparison on the Pulsoid: LNPs were counted particle by particle in three states — freshly thawed, after 80 minutes at room temperature, and after two freeze–thaw cycles. The freshly thawed sample and the one left 80 minutes at room temperature shared one population near 55 nm; after two freeze–thaw cycles, a distinct population of larger particles appeared near 85 nm.
Why do ensemble methods tend to miss this kind of change? Ensemble light-scattering methods report a combined result for all particles and have limited power to resolve subpopulations of similar size, so a small, newly appearing population can be hidden in the overall result. On top of that, the scattering signal rises steeply with size (in the Rayleigh approximation, for particles much smaller than the wavelength, scattered intensity scales roughly with the sixth power of diameter), so an intensity-weighted result reflects which particles scatter more strongly, not how many of each there are. Single-particle counting picks out the new population on its own: under which condition it appears, where it sits and what share of the detected particles it makes up can be read directly from the distribution.
Once the stability readout changes from one average to a full distribution, compare three things at each time point or condition:
- whether the position and width of the main population have changed;
- whether a new population of larger particles has appeared, and what share of the detected particles it represents;
- whether particle concentration and the zeta potential distribution have shifted with it.
The Pulsoid takes a 65 µL load, typically needs 2–5 minutes per sample when several samples are measured in one session (varying with the sample) and measures 30–210 nm; both the ~55 nm and ~85 nm populations in the experiment above fall within that range, which suits repeated measurement of one LNP lot across several time points and treatment conditions.
Sources: IZON Science, Analyse Lipid Nanoparticles with Single-Particle Precision (LNP application page); IZON Science, The Pulsoid (product page).
Putting single-particle data into the workflow

Size, concentration and zeta potential from the same LNP batch need distinct pre-analytical controls and acceptance criteria.
A reproducible TRPS method is usually built in this order: check sample–electrolyte compatibility (particle concentration, salt, viscosity, surfactants and clogging risk); choose the nanopore and stretch range for the expected size and confirm pulse magnitude and event rate with standard particles; establish blanks and background and keep the event rate neither too low nor heavily overlapping; acquire calibrants, samples, process blanks and system-suitability samples in a fixed order; and analyse with predefined thresholds while retaining raw pulses, exclusion rules and calibration information.

Pore selection, calibration, blanks, event rate and thresholds collectively determine reproducibility.
For mRNA-LNPs, fresh, freeze–thawed and stored time-point samples can be placed in one design to compare main-peak shift, tail growth, particle-count change and zeta-potential drift. When dilution is needed, keep the ionic environment as constant as possible and use recovery experiments to confirm that dilution does not trigger dissociation or aggregation.
TRPS also forms orthogonal evidence with other methods: DLS for a quick view of the overall hydrodynamic distribution, NTA for light-scattering-based single-particle size and concentration, chromatography or electrophoresis for lipid and nucleic-acid components, RiboGreen or similar assays for encapsulation efficiency, and functional cell assays for delivery and expression. When methods disagree, first check weighting, detection limits, dilution, refractive index, thresholds and sample change, rather than choosing the result that looks better.

TRPS, composition, encapsulation and functional assays form complementary evidence for particles, cargo and potency.
Exoid or Pulsoid: deep development work vs QC release
Exoid: flexibility and in-depth characterization. Exoid combines TRPS with tunable polymer nanopores, giving researchers a high degree of control and flexibility: size and concentration over 40 nm–11 µm, concentration from 10⁵ to 10¹¹ particles/mL, and compatibility with nanoparticle types from LNPs and exosomes to virus-like particles. It is the main choice for method development and in-depth characterization.
Pulsoid: speed and standardization. Pulsoid uses nanopore pulse sensing (NPS) with precision solid-state silicon nanopore chips for fast, standardized measurement of LNPs and other particles: 2–5 minutes per sample, a measurable size range of 30–210 nm from a 65 µL load, high sizing accuracy, concentrations of 10⁹–10¹² particles/mL, about twice the throughput of conventional TRPS, and 99.7% agreement with TEM — suited to high-frequency, standardized release testing in QC.
| Need | Better-suited platform | Why |
|---|---|---|
| Formulation development and deep analysis of subpopulations and charge | Exoid | Tunable nanopores, 40 nm–11 µm, many particle types |
| Size and zeta-potential comparison of empty vs loaded LNPs | Exoid | Single-particle size and zeta distributions that reveal charge subpopulations |
| High-frequency, standardized QC release | Pulsoid | 2–5 min per sample, about 2× throughput, 99.7% agreement with TEM |
| Storage and freeze–thaw stability, lot-to-lot distribution comparison | Pulsoid | Particle-by-particle distributions show a newly appearing population of larger particles directly; measures 30–210 nm |
One boundary is worth remembering: results depend on the nanopore, electrolyte, calibration particles and event thresholds, so confirm system suitability with the target sample before method transfer or release use.
Let TechWave help you make LNP characterization count
If your LNP data still stop at one average size and one average zeta potential, or you are preparing traceable particle data for lot release, process changes or regulatory filings, contact TechWave. Based on your sample type, concentration range and testing scenario, we can help compare Exoid and Pulsoid, share application materials and arrange a quotation.
Uses and applicable scenarios
- Comparing size and charge subpopulations of empty and mRNA-loaded LNPs
- LNP formulation screening and before/after comparison of process changes
- Tracking aggregates in storage-stability studies and comparing lot consistency
- High-frequency, standardized particle release testing in QC
Limitations
- Results depend on the nanopore, electrolyte, calibration particles and event thresholds; method suitability must be confirmed for the target sample
- Zeta potential changes with pH, ionic strength and buffer composition and cannot alone represent storage stability, delivery efficiency or in-vivo potency
Frequently asked questions
If the average LNP size barely changes after freeze–thaw, is the sample fine?
That conclusion cannot be drawn directly. In IZON's comparison on the Pulsoid, freshly thawed LNPs and those left 80 minutes at room temperature shared one population near 55 nm; after two freeze–thaw cycles a distinct population of larger particles appeared near 85 nm. A single-particle distribution shows a new subpopulation like this directly; the average may not.
Can empty LNPs and mRNA-LNPs be told apart by size?
In IZON's Exoid comparison, the two size distributions could be told apart: for one formulation, empty LNPs had a modal size of 71 nm and mRNA-loaded LNPs 66 nm, a difference of a few nanometres that is clearly visible in the single-particle distribution.
For LNP characterization, how do I choose between the Exoid and the Pulsoid?
Choose the Exoid for formulation development and in-depth work on subpopulations and charge distributions; its tunable nanopore covers 40 nm–11 µm for size and concentration. Choose the Pulsoid for frequent, standardized QC testing and multi-time-point stability comparison: typically 2–5 minutes per sample when several samples are measured in one session, 30–210 nm, and 99.7% sizing agreement with TEM on NIST-traceable standard particles within the validated measurement range.
Why must measurement conditions be fixed when comparing zeta potential across LNP formulations?
Because the measurement environment itself changes the zeta potential reading. Zeta potential describes the potential near the slipping plane and is affected by the medium, pH, ionic strength and temperature; if conditions differ, an apparent change may come from the measurement environment rather than the LNP itself.